• {isset($vo.title)?$vo.title:''}
  • {isset($vo2.title)?$vo2.title:''}

Like

Multi-GAS high-precision UV gas sensors

  • Product modelS-Ultra.Sens
  • Product brand smartGAS
  • Apply area:Medical Equipment,Wearable Devices,Aged Care Facility,Marine Equipment,Petrochemical Industry,Electric Power,Light Industry,Special Industry,Equipment Manufacturing,Inter-city Rail Transport/Metro,Instrumentation and Measurement,Automation Cntrol,semiconductor,Warehouse,Nuclear Power,Metallurgy/Steel Industry,HVAC,Machine Engineering,Telecommunication,Transportation and Road Administration,Internet of things,Smart Phone,Smart Home,Oil-fueled Automotive,Industry Robtot, Water Conservancy and Transportation,New energy(Wind Energy,Photovoltaik,Hydrogen,Geothermy etc.),Commercial Archiecture,Personal Management,Indoor Environment Monitoring,Lighting Management,Energy Management,Health Care Industry,Entertainment Equipment,Personal Computer,Service Robot,Education Robot,Rehabilitation Robot,New Energy Vehicle,High-speed Train,Aviation, Refrigeration,Household Appliance,Food/Beverage/Medicine Industry,Transportation,Retail,Animal Husbandry,Breading Industry,Warm Shed,Agricuhural Machinery,Boats and Ships,Outer Space and Sapce flight,Petroleum/Natural Gas,Fuel Cell,Energy Safty and Measurment,Energy Transportation,Outdoor Environment Monitoring,Security Protection/Monitor,Exploration/Monitoring,Water/Sewage Disposal,Industry Environment Monitoring,Underground Pipline,Smart Water,Water Supply and Drainage,Urban Rainstorm Waterlogging Management

smartGAS Sensor Technology Co.,Ltd.

  

Product Details Product specification download

Products Description:

 

The gas analysis based on the NDUV techniques is a well-known method to determine the concentration in complex gas mixtures. The S-Ultra.Ssens module uses modern optical components to achieve optimized analyzer results.

 

We use a special LED/EDL-Technology in order to get stable radiation emission. The result is a high resolution gas analysis system which can detect gas concentration changes of less than 1ppm .

 

The entire structure can be disassembled to make the maintenance/service easier. The different devices are sealed with O-Ring joining. The stainless steel sample cell is coated with a protection layer. The length is 100mm. For the combined detection of NO, NO2 and SO2 three different light sources are integrated.

 

For example, we use AlGaN-LED for the UV light source which can cover 360-230nm wave length. For enhance the life time of the light source, we use special control board, which can let the light source achieve more than 3 years life time, which solve the well-know problem of shortcoming of UV light source.  

 

S-Ultra.Sense is suitable for all kind of complex industry environment, special has good performance in low range CEMS measurement, which can good measure NO and NO2 without transition device.

 

Standard gas measurement type (other gas and range please contact us):

 

S-UltraSense

NO

0-300ppm

S-UltraSense

O3

0-100ppm

S-UltraSense

O3+CnHm

O3 0-25ppm  /  CnHm 0-500ppm

S-UltraSense

H2S

H2S 0-100ppm / 0-500ppm / 0-1500ppm / 0-5000ppm

S-UltraSense

H2S (Background gas CH4)

H2S 0-100ppm

S-UltraSense

NO/NO2/SO2

NO 0-300ppm / NO2 0-200ppm / SO2 0-200ppm

S-UltraSense

SO2/NO2

SO2 0-1000ppm / NO2 0-1000ppm

S-UltraSense

NO2/N2O

NO2 0-2000ppm / N2O 0-2000ppm

S-UltraSense

N2O

0-1000ppm

S-UltraSense

SO2

0-100ppm / 0-500ppm

S-UltraSense

SO2 dual range

0-1000ppm (LDL 1ppm), 0-30000ppm(LDL 20ppm)

 

Product Advantages:

Fast readout time 200ms

 High dynamic range

 No cross talk to H2O

 Rugged sensor design

 Demountable construction

 Gas tight O-Ring sealing

 Low power consumption <1 Watt @ 24 VDC

 Different Interfaces (RS 232, CANopen)

 Connection of two optical benches 


Specification:

 

Measurement Technology

NDUV

Heating Tem.

42.10

T90 Response time

 <3s@1L/min

Digital resolution@zero)

1ppm...0.1 Vol.%

Detection Limit (3sigma)

±0.5%FS

Repeatability

±1%FS

Linearity Error

±1%FS

Stability (zero)

±1%FS

Stability (span)

±1%FS

Temperature Drift. (zero)

±1%FS per 10K(典型值)

Temperature Drift. (span)

±1%FS per 10K(典型值)

Pressure Drift.

±0.015% /mbar

Operating Temp

10...40℃

Storage Temp  

-10...60℃

Ambient Humidity  

0 % up to 95 % rel. Humidity (not condensing )

Ambient Air Pressure

800...1200 mbar

Warm up Time

<60min (related to environment Tem.)

<30min @20℃ environment Tem.

Start time

<2 min (reach the heating Tem.)

Flow gas Tem.

-2℃...40℃

Flow gas Humidity

Max. 10℃ (not condensing)

Gas flow

0.2...1L/min (stable flow)

Output Signal

RS232 9P-Sub-D 和 2x5pin 2.45mm 接头

Operating Voltage

24VDC±0.3V 3A

Power consumption

4W (Heating max. 55W)

 


© Copyright © 2021 Shanghai Technology Convention & Exhibition Co. LTD. All rights reserved. 沪ICP备13047538号